A stylus that's too long or too heavy can still "work," but the results quietly get worse, and on some probes it can stop working right. This article covers the limits for the VAST XXT and VAST XT and how to pick a stylus system that stays inside them.

Every stylus bends a little when it touches the part. A longer, heavier, or more jointed stylus bends more, and that bending shows up as measurement error. On the VAST XXT the stylus system is held on magnetically, so too much weight, or weight sitting off to one side, can pull or tilt it.
ZEISS describes the ideal stylus system as one that:
<aside> 💡
Key tip: Use the shortest and lightest stylus that can reach the feature. Every extra joint, millimeter, and gram costs accuracy. Use the largest shank diameter to reduce stylus bending.
</aside>
Weights include the adapter plate.
| Probe | Max stylus length | Max weight | Notes |
|---|---|---|---|
| VAST XXT TL1 | 125 mm | 10 g | Side styli: max 40 mm horizontally, at least 50 mm from the adapter plate in the Z direction |
| VAST XXT TL2 | 250 mm | 10 g | Has a fixed 75 mm extension. Side styli: max 40 mm horizontally |
| VAST XXT TL3 | 150 mm | 15 g | Side styli: max 65 mm horizontally, at least 50 mm from the adapter plate in the Z direction |
| VAST XXT TL4 | 250 mm (vertical) | 10 g | Built for long vertical reach. Side length max 40 mm horizontally |
| VAST XT | up to 500 mm (including extensions) | 500 g | Max imbalance torque 0.3 Nm |
| VAST XTR Gold | 350 mm | 500 g | Max imbalance torque 0.15 Nm |
On the VAST XXT, you can also add a probe extension of up to 100 mm between the RDS-XXT adapter plate and the VAST XXT probe (TL1, TL2, and TL3).